Payment Terms | T/T(wire transfer), Western Union, Paypal |
Delivery Time | 3-5 working days |
Packaging Details | Carton |
Product Name | Mindray IMEC10 Patient Monitor Power Supply Board KB26Q5463 009-002108-00-2.0 |
Brand | Mindray |
Part No | KB26Q5463 009-002108-00-2.0 |
Use for | Mindray IMEC10 Patient Monitor |
MOQ | One piece |
Warranty | 90 days |
Line voltage | 100 to 240 VAC (±10%) |
Current | 1.1 to 0.5 A |
Frequency | 50/60 Hz (±3Hz) |
Condition | Work in Good |
Brand Name | Mindray |
Model Number | KB26Q5463 009-002108-00-2.0 |
View Detail Information
Explore similar products
Patient Monitor Accessories Goldway UT4000 Monitor Blood Pressure Plate Great
Mindray PM7000 Patient Monitor Accessories Keyboard Silicone 7000-20-24395
Well-chosen Patient Monitor Mainboard Goldway G30 40 60 70 80 UT4000B UT6000
Patient Monitor Repair Parts GE Dash4000 Patient Monitor Flat Cable in ready
Product Specification
Payment Terms | T/T(wire transfer), Western Union, Paypal | Delivery Time | 3-5 working days |
Packaging Details | Carton | Product Name | Mindray IMEC10 Patient Monitor Power Supply Board KB26Q5463 009-002108-00-2.0 |
Brand | Mindray | Part No | KB26Q5463 009-002108-00-2.0 |
Use for | Mindray IMEC10 Patient Monitor | MOQ | One piece |
Warranty | 90 days | Line voltage | 100 to 240 VAC (±10%) |
Current | 1.1 to 0.5 A | Frequency | 50/60 Hz (±3Hz) |
Condition | Work in Good | Brand Name | Mindray |
Model Number | KB26Q5463 009-002108-00-2.0 | ||
High Light | Mindray IMEC10 Patient Monitor Parts ,KB26Q5463 Patient Monitor Parts ,009-002108-00-2.0 Power Supply Board |
Mindray IMEC10 Patient Monitor Power Supply Board KB26Q5463 009-002108-00-2.0
AC and DC power supply board
The power supply board is located on a metal sheet in the battery compartment on the bottom of the monitor. The power supply converts incoming AC power (100-240 VAC, 50/60 Hz) to a VIN bus voltage of 15.2 V and serves as the power source for all monitor voltages. The AC/DC power supply board supplies power to all modules via a 2-pin power cable on the connection board.
Connection card (DC/DC, power management circuit)
Breakout board for sending signals from external connectors and DC/DC power management circuits
Power Supply Specifications
AC power | |
Line voltage | 100 to 240 VAC (±10%) |
Current | 1.1 to 0.5 A |
Frequency | 50/60 Hz (±3Hz) |
Symptoms | Possible Cause | Corrective Action |
The patient monitor fails to start: AC LED, power on/off LED or battery LED does not light
| AC mains not connected or battery too low
| Check that AC mains is properly connected or battery capacity is sufficient
|
Cables defective or to start: AC poorly connected
| 1. Check that the cables from power switch & LED board to button board, button board to main board, power board to interface board, and interface board to main board are correctly connected. 2. Check that cables and connectors are not damaged
| |
Power switch & LED board defective | Replace the power switch & LED board | |
Power board defective | Replace the power board | |
Interface board defective | Replace the interface board. | |
Mother board Defective | Replace the mother board |
1. Disconnect the cable for AC input receptacle. Remove the three M3X6 screws indicated below and remove the power board
NOTE
Follow the instructions and unscrew the correct screws.
Thank you so much for your attention .If you're interested in Mindray IMEC10 Patient Monitor Power Supply Board KB26Q5463 009-002108-00-2.0 ,feel free to contact us .Thank you .
Company Details
Business Type:
Manufacturer,Distributor/Wholesaler,Agent,Importer,Exporter,Trading Company,Seller
Year Established:
2015
Employee Number:
20~50
Ecer Certification:
Verified Supplier
China World Technology Medical Equipment Service Group is a company engaged in medical equipments/repair parts and accessories products trading company. The company was established in September 2005, has many famous enterprises at home and abroad (such as America USCO company) to establish a long-te... China World Technology Medical Equipment Service Group is a company engaged in medical equipments/repair parts and accessories products trading company. The company was established in September 2005, has many famous enterprises at home and abroad (such as America USCO company) to establish a long-te...
Get in touch with us
Leave a Message, we will call you back quickly!